Experiment · Medical imaging

I rebuilt my chest X-ray model: 0.80 AUROC on a patient-wise split, and probabilities up to 11 times too high

My first model reported 58% accuracy. The rebuilt version uses the right task, a patient-wise split, and checks whether its probabilities can be trusted.

By Muzammal Hussain · 29 October 2026 · 4 min read

Update, 30 October: a follow-up run with three seeds, a validation-fitted correction and subgroup results is in the second rebuild.

My first version classified chest X-rays from the NIH ChestX-ray14 dataset and reported about 58% accuracy. Looking back, that number could not be trusted: the task was set up wrongly, the metric hid rare diseases, and the same patient could appear in training and validation. I rebuilt it in one Kaggle notebook and then checked something the first version never looked at: whether the model's percentages mean what they say.

Key findings

  • With 14 labels, the official patient-wise split and per-class AUROC, DenseNet121 reached a test mean AUROC of 0.801 (95% interval 0.796 to 0.806) and ResNet50 0.795, on 25,596 test images.
  • DenseNet121 was ahead in a paired bootstrap by 0.006 (interval 0.003 to 0.010). Small, but it did not flip in 1,000 resamples.
  • The raw probabilities were far too high: 2.5 to 10.9 times the real share of images with each label. For consolidation the model said 49% on average, while 7% of test images had it.
  • A simple per-class correction (Platt scaling) cut the mean calibration error from 0.265 to 0.005 without changing the AUROC.

What changed from the first version?

V1V2
Task836 label combinations as classes14 labels, one sigmoid output each
Splitrandom by imageofficial split, by patient
Metricaccuracyper-class AUROC
Result58.1% validation accuracy0.801 test mean AUROC (DenseNet121)

The two results measure different things, so they cannot be compared. The V2 split puts 77,719 images in training, 8,805 in validation and 25,596 in test, and the code checks that no patient appears in two of them. Both models were ImageNet-pretrained, trained at 224x224 with the same settings on a Kaggle T4 GPU, and stopped after 7 epochs, keeping the weights from epoch 5. After that the training loss kept falling while the validation AUROC dropped, which is the overfitting V1 never caught.

Horizontal bar chart of test AUROC for 14 chest X-ray labels for DenseNet121 and ResNet50. Highest are emphysema, hernia and cardiomegaly near 0.88 to 0.91; lowest are infiltration, pneumonia and consolidation near 0.69 to 0.73.
Figure 1. Per-class test AUROC on the official patient-wise test split. One run per model.

The easiest labels were emphysema (0.912), hernia (0.892) and cardiomegaly (0.879). The hardest were infiltration (0.695), pneumonia (0.711) and consolidation (0.728). On an X-ray all three appear as white opacity where air should be, so they overlap visually, and their labels, which were mined from report text, may also be the least reliable.

The percentages were wrong

AUROC only asks whether a sick patient gets a higher score than a healthy one. It says nothing about whether "49%" means 49%. I compared the average predicted probability with the real share of test images for each label.

Bar chart comparing, for 14 labels, the share of test images with the label and DenseNet121's mean predicted probability. The predicted probability is several times higher for every label.
Figure 2. Mean predicted probability against the real share of test images with each label, DenseNet121.

Every label was overestimated, from 2.5 times for infiltration to 10.9 times for pneumonia. The likely cause is a choice I made on purpose: during training, rare labels were given more weight so the model would not ignore them. That helps the ranking but inflates the probabilities. Fitting a one-line logistic correction for each label on half of the test images and applying it to the other half brought the mean expected calibration error from 0.265 to 0.005 and the mean Brier score from 0.175 to 0.058. The AUROC stayed at 0.801.

I also tried letting the model refer its least certain images to a person. Keeping the 50% of images it was most sure about raised the mean AUROC on those images from 0.801 to 0.816. This used the uncorrected probabilities, so it is a first look only.

What I noticed

Limitations

Reproduce the experiment

The repository has the Kaggle notebook, the saved test predictions, the analysis script and its unit tests. The analysis runs on a Mac with Python 3.9 in a few minutes: python3 reliability_analysis.py --results results --out analysis. Training both models takes under two hours on a free Kaggle T4.

Data: NIH ChestX-ray14 (Wang et al., 2017), from the Kaggle dataset nih-chest-xrays/data. Related reading: Rafferty and Rajan, 2025, on the limits of public chest X-ray datasets. This is a small engineering experiment, not a peer-reviewed paper.